Evaluating dosimetric parameters with a plastic scintillator for megavoltage photon beam quality assurance.

Created on 11 Oct 2025

Authors

Gregory Penoncello, Bernard L Jones, David A P Dunkerley, Adam Mahl, Moyed Miften, Cem Altunbas, C-K Chris Wang, Daniel G Robertson

Published in

Journal of applied clinical medical physics. Volume 26. Issue 10. Pages e70276.

Abstract

Linear accelerators require a large amount of data to be collected daily, monthly and annually to verify safe deliveries for patients. Different detectors have been utilized to improve the simplicity, efficiency and accuracy of the various experimental setups required to collect the necessary data resulting in reduced data collection and evaluation time. Plastic scintillators are stable and energy independent radiation detectors with high spatial resolution that emit light proportional to the amount of radiation incident on them. With an appropriate photodetector and technique to measure and analyze the light that is emitted, scintillators can be utilized to measure dosimetric parameters necessary for various monthly quality assurance requirements.
A uniform cylindrical plastic scintillator imaged by a complementary metal oxide semiconductor (CMOS) camera is designed to act as a radiation detector to measure 2D projections of linear accelerator beams. Beams were delivered to the detector to evaluate machine quality assurance (QA) parameters, including beam energy, output, profile consistency and reproducibility as described by TG-198. These measurements were compared to calculations from the Eclipse treatment planning system (TPS). Typical monthly quality assurance (QA) beams were delivered and 2D projections of the scintillation light were measured to validate the accuracy and reproducibility of this detector system for monthly QA.
The plastic scintillator was able to accurately characterize the radiation beam. Energy and profile measurements were reproducible and within 2%/2mm of calculations in the TPS. Output measurements had maximum variations of up to 1.3% and average differences of 0.5%.
A simple cylindrical plastic scintillator and CMOS camera radiation detector setup was designed and tested for measuring monthly QA dosimetric parameters specified by TG-198 with accurate and reproducible output, energy and profiles measurements. This method reduces the number of measurements required, allowing multiple parameters to be evaluated in a single beam delivery.

PMID:
41074225
Bibliographic data and abstract were imported from PubMed on 11 Oct 2025.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 73
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement