[Reliability Analysis of Wide-Detector CT Equipment Based on Improved FMEA].

Created on 24 Jun 2026

Authors

Wenhao Xu, Wei Jin, Shundong Hu

Published in

Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation. Volume 50. Issue 3. Pages 306-310. May 30, 2026.

Abstract

An improved failure mode and effects analysis (FMEA) model is constructed to enhance the accuracy of reliability assessment for wide-detector computed tomography equipment.
A multidisciplinary expert panel is established. A scoring system is used to quantitatively evaluate various potential failure modes, and expert weights are introduced for weighted calculation to achieve objective ranking of risk priority.
The overheating of the X-ray tube and system communication interruption are identified as the highest-risk failure modes. Significant differences in equipment reliability under different clinical usage patterns are quantitatively revealed.
The improved FMEA method can effectively identify key failure links, providing a reliable basis for formulating differentiated and precise preventive maintenance strategies, which is helpful for ensuring the continuous and stable operation of equipment.

PMID:
42338371
Bibliographic data and abstract were imported from PubMed on 24 Jun 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 8
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement