Sivan Unit (SU) for standardization of voxel intensity in CBCT: A multi-vendor, multi-centric validation study.

Created on 30 Jun 2026

Authors

Sivan Sathish

Published in

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). Volume 148. Pages 105868. Jun 29, 2026. Epub Jun 29, 2026.

Abstract

To develop and validate the Sivan Unit (SU), a biologically referenced voxel intensity harmonization framework for cone-beam computed tomography (CBCT) designed to improve cross-vendor grayscale consistency in quantitative maxillofacial imaging.
This retrospective multi-vendor, multi-centric study analyzed 160 CBCT datasets from ten commercial scanner systems. Raw voxel intensities (VIraw) from seven maxillofacial tissues were first processed using global Z-score transformation to reduce gross scaling differences. These were subsequently mapped using two internal biological anchors: dental pulp (0 SU) and cortical bone (1000 SU). Inter-scanner consistency was evaluated using one-way ANOVA, coefficient of variation (CV), intraclass correlation coefficients (ICC), and Bland-Altman analysis.
Pre-harmonization intensities demonstrated significant inter-scanner variability across all tissues (p < 0.001), with a mean CV of 14.7%. Following SU transformation, voxel intensity distributions showed no significant inter-vendor differences (all p > 0.05), with the mean CV reduced to 2.1%. Reliability was excellent across all categories (ICC: 0.89-0.97). Bland-Altman analysis confirmed a marked reduction in inter-scanner bias and a narrowing of 95% limits of agreement. The intrinsic biological hierarchy (enamel > cortical bone > dentin > trabecular bone > pulp > soft tissue > air) was preserved post harmonization (r = 0.992, p < 0.001).
The Sivan Unit (SU) provides a biologically anchored harmonization framework that effectively eliminates vendor-dependent variability without external phantoms or absolute physical calibration. By ensuring cross-platform reproducibility, the SU framework provides a standardized quantitative foundation that may support future radiomics and AI-based applications in maxillofacial radiology.

PMID:
42372359
Bibliographic data and abstract were imported from PubMed on 30 Jun 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 2
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement