Intrasaccular Flow Disruption with the Contour Device: A Single-Center Experience in Ruptured Intracranial Aneurysms.

Created on 21 Jul 2026

Authors

Deepak K Singh, Praveenkumar N Sangolli, Prevesh Sharma, Vipin Chand, Vineet K Mishra, Neha Singh

Published in

Neurology India. Volume 74. Issue 4. Pages 629-636. Jul 01, 2026. Epub Jul 21, 2026.

Abstract

Management of wide-necked intracranial aneurysms remains a challenge for clinicians. Intrasaccular devices like Contour and WEB (Woven Endo Bridge) devices offer new promise in managing wide-necked intracranial aneurysms. The Contour Neurovascular System, developed by CERUS Endovascular (Fremont, California, USA), is an intrasaccular device working on the principle of flow-disruption and flow diversion. We report our institutional experience with this device in cases of ruptured intracranial aneurysms.
To investigate the role of the Contour device as a first-line intervention in ruptured intracranial aneurysms, focusing on its safety, efficacy, and patient outcomes.
This is a retrospective study conducted in the Department of Neurosurgery at Dr. Ram Manohar Lohia Institute of Medical Sciences (RMLIMS), Lucknow, India. A single surgeon operated on all cases. Clinical, radiological, and follow-up data were collected and analyzed for all patients treated with the Contour device. Modified Rankin scale (MRS) was used at 6 weeks to assess the patient's performance. MR angiography (MRA) brain was done at follow-up to assess the aneurysmal occlusion using the Raymond-Roy classification (RROC).
Of the 15 cases of treated ruptured intracranial aneurysms, 10 were anterior communicating artery (ACom) aneurysms, 3 were middle cerebral artery (MCA) bifurcation aneurysms, and 2 were basilar tip aneurysms. The contour device was successfully deployed in all the cases. At 6-month follow-up, MRA revealed that 11 patients had Raymond-Roy Occlusion Classification (RROC) Class I occlusion, 2 patients had neck remnants (RROC Class II), and 1 patient had RROC Class IIIa occlusion with partial aneurysmal reperfusion.
Intrasaccular flow disruption is a novel endovascular approach for the treatment of wide necked intracranial aneurysms using devices like Contour. The Contour device demonstrated a favorable safety profile and proved to be a stable and effective treatment option, even in the acute setting of aneurysmal rupture.

PMID:
42478396
Bibliographic data and abstract were imported from PubMed on 21 Jul 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 10
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement