Helium pencil beam commissioning and beam modeling.

Created on 08 Aug 2026

Authors

Lukas Martin, Barbara Knäusl, Peter Kuess, Dietmar Georg, Hugo Palmans, Lorenz Wolf, Nadia Gambino, Fabio Farinon, Andrej Prochazka, Mansure Schafasand, Lars Glimelius, Walter Ikegami Andersson, Daniel Simon Colomar, Antonio Carlino, Markus Stock, Hermann Fuchs

Published in

Physics and imaging in radiation oncology. Volume 40. Pages 101046. Epub Jul 26, 2026.

Abstract

Helium ions combine reduced lateral scattering compared to protons and a lower fragmentation tail than carbon ions, enabling sharp dose gradients and improved normal tissue sparing. This study reports the commissioning of a scanned helium pencil beam line and validation of its corresponding beam model.
Synchrotron-based helium ion beams were commissioned covering energies from 54.6 to 402.8 MeV/u. Depth-dose curves were measured and absolute dose calibration was performed. Beam optics (spot size, position, and intraspill stability) were evaluated for various spill lengths. A beam model was implemented in the RayStation treatment planning system (TPS) and validated through 2D absolute dose measurements in homogeneous and heterogeneous phantoms and 3D measurements of cubic spread-out Bragg peak fields. Gamma-index analysis and Monte Carlo (MC) simulations with GATE/Geant4 were performed for benchmarking.
Measured ranges agreed with MC simulations within ± 0.3 mm. Spot sizes decreased with energy, independently of the spill length. Spot positions remained within ± 0.5 mm and intraspill variations were 0.2 mm (position) and 5.4 % (size). TPS-predicted doses agreed within 0.1 %. For 3D validations in homogeneous phantoms, the dose differences were generally within 2 %. Median gamma pass rates exceeded 90 % for 3 %/1.5 mm and 95 % for 5 %/1.5 mm. For the heterogeneous phantom, differences were within -3.2 %.
Stable scanned beam delivery with helium ions was established. Validation demonstrated strong agreement between measurements, TPS calculations, and MC simulations, supporting research and future clinical application.

PMID:
42569120
Bibliographic data and abstract were imported from PubMed on 08 Aug 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 5
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement