Authors
Nicolás Padilla-Raygoza, Gilberto Flores-Vargas, María de Jesús Gallardo-Luna, Efraín Navarro-Olivos, Luis Guillermo Patiño-Gutiérrez, Ariana Valeria Girón-Soto, Evaristo Antonio Meza-Galván, Itzel Marcela Anguiano-Canchola, Sergio Emmanuel Luna-Santillán, José Juan Torres-Hernández
Published in
Frontiers in pediatrics. Volume 14. Pages 1801057. Epub Jul 27, 2026.
Abstract
Developmental dysplastic hip disease is common in childhood, and clinical maneuvers (e.g., Ortolani and Barlow) detect only subluxation or dislocation. Sound transmission has been proven useful in the diagnosis of this disease.
to calculate the validity and repeatability of electromagnetic acoustic wave transmitter-receiver device for the clinical diagnosis of hip dysplasia in newborns.
This was a cross-sectional study of newborns at Guanajuato General Hospital whose parents agreed to participate.
One hundred newborns, between 4 and 27 days of extrauterine life, underwent sound transmission testing with the punctual sound emitter-receiver device placed separately on each hip on three occasions by two examiners. Hip ultrasound was performed using the Graf technique as the gold standard. Validity (sensitivity, specificity, and predictive values) and Cohen's Kappa were calculated to determine the reliability of the device.
Of the 200 hips, 15 right and 18 left hips were diagnosed with dysplasia using the Graf technique ultrasound; a validity greater than 90% was obtained in the three measurements. Repeatability with intra-observer Kappa values were 0.76 for the right hip and 0.82 for the left hip, while inter-observer Kappa values were 0.90 for the right hip and 0.81 for the left hip.
Sound transmission using this clinical device complements the evaluation of the hips of newborns.
PMID:
42577330
Bibliographic data and abstract were imported from PubMed on 11 Aug 2026.