Comparison of Four CBCT-IGRT Geometric QA Phantoms within a Single Institutional Workflow.

Created on 21 Aug 2026

Authors

Hyojun Park, Hyejo Ryu, Jin Hwa Choi, So-Yeon Park, Yoonsuk Huh, Seonghee Kang, Chang Heon Choi, Jung-In Kim, Minsoo Chun

Published in

The British journal of radiology. Aug 21, 2026. Epub Aug 21, 2026.

Abstract

To compare four commercial cone-beam CT (CBCT) geometric quality assurance phantoms within a common workflow, focusing on positioning/repositioning accuracy and agreement between the laser-defined setup reference and the CBCT imaging isocentre.
MIMI™, QUASAR™ Penta-Guide, ISOCube, and MZ CUBE were evaluated monthly for five months on a Versa HD system with XVI CBCT and a HexaPOD couch. Two observers independently completed each measurement. Three-dimensional post-correction translational residuals (Δdpos/rep) and laser-defined setup reference-CBCT imaging isocentre distances (Δdiso) were calculated. Phantom comparisons used observer-averaged values from the five monthly sessions. Rotational correction and CBCT image quality were assessed descriptively with MZ CUBE.
Mean Δdpos/rep values ranged from 0.19 to 0.31 mm, with mean absolute observer differences of 0.05-0.12 mm. No overall difference was found among the four phantoms (p = 0.547). Mean Δdiso values ranged from 0.68 to 0.87 mm, and all measurements were below 2 mm. An overall difference was observed for Δdiso (p = 0.019), although no pairwise comparisons were performed. With MZ CUBE, mean post-correction translational and rotational residuals were 0.24 mm and 0.37°, respectively, and image-quality measures showed no consistent temporal trend.
The four phantoms showed similarly small post-correction translational residuals. In our setting, MZ CUBE additionally supported rotational and image-quality checks.
Direct comparison within a common workflow showed similar translational performance, while the scope of additional testing depended on the phantom configuration and locally available accessories.

PMID:
42627341
Bibliographic data and abstract were imported from PubMed on 21 Aug 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 5
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement