Authors
Chao Jing, Kun Liu, Guowei Zhang, Hongwei Wu, Zhe Jia, Guoxin Li, Yaping Zhao, Zhongtao Zhang, Wei Guo
Published in
The international journal of medical robotics + computer assisted surgery : MRCAS. Volume 22. Issue 5. Pages e70221.
Abstract
Single-port robotic cholecystectomy offers minimally invasive benefits. This trial evaluated the feasibility and safety of a novel system, MicroPort SA-1000.
In this dual centre, single-arm, prospective study, eligible patients underwent single-port robot-assisted cholecystectomy with MicroPort SA-1000. The primary endpoint was the completion rate, while secondary endpoints included estimated blood loss, operating time, hospital length of stay and complications.
From June 2023 to December 2023, a total of 44 patients from two centres were enrolled, and all patients underwent single-port robot-assisted cholecystectomy without conversion. The median EBL was 10 mL (range 5-200 mL). The total operative time was 88.43 ± 25.87 minutes. The median LOS was 2.5 days (range 1-7 days). No serious surgery-related adverse events were observed.
The novel single-port robot demonstrated feasibility and safety in cholecystectomy with effective triangulation, controlled operative time and reduced trauma. Larger studies are needed to validate its performance across different surgical fields.
This study was registered at https://www.chictr.org.cn/showproj.html?proj=258677 (ChiCTR2500107401).
PMID:
42667626
Bibliographic data and abstract were imported from PubMed on 30 Aug 2026.