Impact of collimator misalignments on dosimetric robustness in proton minibeam radiotherapy using an automated measurement setup.

Created on 07 Sep 2026

Authors

Sake Salverda, Ernst van der Wal, Thomas Toet, Mischa Hoogeman, Kelvin Ng Wei Siang

Published in

Medical physics. Volume 53. Issue 9. Pages e70671.

Abstract

Proton minibeam radiation therapy is an emerging technique that combines the tissue-sparing benefits of protons with the spatial modulation used for minibeams to reduce toxicity in healthy tissue. Achieving accurate delivery of planned dose distributions depends on precise collimator-to-beam alignment, for which it is essential to understand the behavior of both multi-slit and single-slit configurations. Understanding single-slit behavior is particularly important, as it also linearly affects the heterogeneity (peaks and valleys) of the overall minibeam. While several studies have investigated misalignment effects with multi-slit collimators, there is limited experimental characterisation of single-slit behavior and its impact on multi-slit arrays.
This study aims to quantify the dosimetric effects of translational and rotational collimator-to-beam misalignments for single-slit proton minibeams using our in-house automated measurement setup. Additionally, the robustness of lateral dose profiles for minibeam arrays will be investigated under varying collimator misalignments.
A single slit collimator with 150 M e V protons was used to produce a single minibeam with adjustable collimator slit widths (0.4 and 0.7 m m ). We introduced rotational and translational misalignments using motorized stages. We developed a quick measurement setup to measure lateral profiles using a microDiamond detector. The lateral profiles were measured at depths of 1 and 13 c m using a water-equivalent phantom. We validated the setup by comparing the measurements to gafchromic film measurements using local gamma analysis. A custom fitting model combining two Gaussians and a Lorentzian was developed to fit the lateral profiles. From the fitted profiles, we determined the maximum dose, the lateral position of maximum dose, the full width at half maximum, and the symmetry. Minibeam arrays were simulated via superposition of single-slit profiles. Specific misalignment combinations were measured for 140, 160, and 170 M e V protons, energies typically used in the clinic for deep seated tumours, to assess how misalignments affect the minibeam at different energies.
Rotational misalignments caused substantial asymmetry and peak dose reductions up to 68.4% at shallow depths, with asymmetry reducing at greater depths. Translational shifts resulted in a translation of the minibeam profile in the same direction and magnitude without changing the minibeams shape. Despite these distortions, simulated minibeam arrays achieved acceptable dose homogeneity at 13 c m depth for central 80% field-widths with flatness below a 3% threshold. The proposed measurement setup enabled profile acquisition of a single minibeam in under one minute with submillimetre resolution and strong agreement with gafchromic film (3%/ 0.2 m m gamma pass rates of 100%).
Collimator misalignments substantially affect the minibeam dose robustness and should be considered in clinical treatment planning and quality assurance (QA). The presented measurement setup enables, rapid, high-resolution profile measurements using commercially available detectors and can facilitate routine clinical QA.

PMID:
42703749
Bibliographic data and abstract were imported from PubMed on 07 Sep 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 9
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement