Authors
Euntaek Yoon, Sung Hyun Lee, Hyeongmin Jin, Tae Ho Jang, Sungjin Yoon, Hyung Jin Choun, Soon Ho Yoon, Jin Mo Goo, Jong Min Park
Published in
PloS one. Volume 21. Issue 9. Pages e0358184. Epub Sep 15, 2026.
Abstract
To evaluate the impact of stopping-power ratio (SPR) predictions derived from dual-layer detector-based spectral CT (DLCT) on conventional single-energy CT (SECT)-based carbon-ion treatment planning for thoracic cancer.
Twenty patients with thoracic cancer who underwent non-contrast DLCT were retrospectively included. DLCT-based SPR images were generated from effective atomic number and relative electron density maps using the Bethe equation. SECT-based SPR was derived using a Hounsfield unit look-up table generated via stoichiometric calibration of a commercial electron density phantom. Using the same phantom, a phantom-based prediction assessment was performed by comparing SECT- and DLCT-based SPR predictions with theoretical SPR values for the inserts. For patient analysis, normal-tissue SPR values derived from SECT and DLCT were compared. Carbon-ion treatment plans were first generated based on SECT images and then recalculated using DLCT-derived SPR images. Range differences in the beam's-eye view and dose-volume histogram parameters for the planning target volume and organs at risk were evaluated.
In the phantom-based assessment, the root-mean-square error relative to the theoretical SPR was 5.57% for SECT and 2.83% for DLCT. In patient images, mean SPR differences between SECT and DLCT were 3.62% for the lungs, 1.93% for the esophagus, 0.04% for the heart, 0.93% for the spinal cord, and 1.53% for the ribs. Across the patient cohort, the mean recalculated range difference was 0.71 ± 0.89 mm, corresponding to a relative difference of 0.60 ± 0.67%. No statistically significant differences were observed in target dose metrics, whereas selected organ-at-risk metrics showed statistically significant differences between SECT-based plans and DLCT-based recalculations.
DLCT-derived SPR prediction showed better agreement with theoretical phantom SPR values in the phantom-based prediction assessment and produced tissue-dependent SPR differences in thoracic patient images compared with conventional SECT-based prediction. Recalculation on DLCT-derived SPR maps resulted in modest but systematic range differences and limited changes in selected dose metrics. These results may inform the implementation and future independent validation of patient-specific DLCT-based SPR workflows in carbon-ion treatment planning for thoracic cancer.