Authors
Varun Kant Yadav, R K Bhatia, A Singh, Rabi Datta, M M Gulhane, E Ravisankar, R D Phulsundar, K G Bhushan
Published in
Rapid communications in mass spectrometry : RCM. Volume 40. Issue 24. Pages e70184. Dec 30, 2026.
Abstract
Thermal ionization mass spectrometry (TIMS) integrated with a high-efficiency cavity ion source (CIS) has emerged as a promising analytical tool for precise isotope ratio measurements and trace elemental analysis in nuclear science. Operation with high ionization potential (HIP) elements such as uranium requires higher cavity temperatures and emission currents. This caused glow discharge, thermal instability, localized overheating, and degradation of ion beam stability during prolonged operation. This necessitates a detailed approach for further enhancement of source performance and operational stability.
The CIS was systematically redesigned for HIP elements through detailed engineering optimization of the cavity geometry, thermionic filament dimension, ion extraction configuration, and shielding electrode design to achieve stable ion beam generation. The optimized CIS assembly was then integrated with indigenous magnetic sector TIMS platform. After that, its analytical performance was systematically evaluated relative to a conventional thermal ionization source (TIS) configuration operated under same experimental conditions. Comparative measurement was carried out in terms of analytical sensitivity (the total detected ion signal per unit sample loading). Also, the time-dependent instrumental mass fractionation behavior and long-term ion beam stability study were carried out.
The design improvements and systematic optimization of the CIS showed nearly an order of magnitude increase in analytical sensitivity for uranium together with a ~1%-2% improvement in mass fractionation stability relative to the conventional TIS configuration.
The design improvements carried out in the CIS configuration resulted in higher analytical sensitivity and better resistance to systematic isotopic drift during uranium measurements. This established the CIS TIMS platform as an effective and high-performance analytical tool. It useful for precise actinide isotope ratio measurements relevant to nuclear safeguards, nuclear forensics, and trace isotopic analysis.
PMID:
42813959
Bibliographic data and abstract were imported from PubMed on 30 Sep 2026.