Evaluation of registration accuracy using short-arc cone-beam computed tomography for image-guided radiotherapy in breast cancer.

Created on 07 Oct 2026

Authors

Shuhei Wada, Masahide Tominaga, Motoharu Sasaki, Koichi Yamanashi, Yoshio Iwai, Shunsuke Furutani, Yusei Shinohara, Yoshihiro Uto

Published in

Journal of applied clinical medical physics. Volume 27. Issue 10. Pages e70804.

Abstract

Image-guided radiotherapy (IGRT) using cone-beam computed tomography (CBCT) enables highly accurate imaging registration; however, increased radiation exposure remains a concern. Recently, short-arc CBCT (SA-CBCT), which reduces projection data and reconstructs images, has become available. Although previous studies have shown that image quality deteriorates when projection data are reduced, no study has directly evaluated the clinical usefulness of SA-CBCT or its effect on registration accuracy.
We verified the registration accuracy of SA-CBCT in breast tangential irradiation using an anthropomorphic phantom and clinical data, and examined its clinical feasibility compared with the commonly used 200° scan.
An anthropomorphic phantom was placed on the treatment bed, and CT and CBCT images acquired during treatment planning were aligned. Bed movement was measured for both conventional CBCT and SA-CBCT images. The difference between the two measurements was quantified as N, defined as the norm of the vector difference between CBCT and SA-CBCT registration shifts, to evaluate registration accuracy. The acquisition angle was varied to assess the effects of central and total angles. In the clinical study, registration data from 30 consecutive patients who underwent whole-breast irradiation were analyzed. Data were reconstructed by changing the acquisition angle from 200° to 120°, and N was calculated as in the phantom experiment.
In the phantom experiment, N was within 0.15 mm (3-axis) and 0.55 mm (6-axis) when the acquisition angle was reduced to 120°. In clinical data, the average N was 0.64 mm in 3-axis registration (84.7% within 1 mm) and 0.85 mm in 6-axis registration (71.3% within 1 mm). Even with a 120° acquisition angle, the body contour and chest wall edges were sufficiently preserved.
SA-CBCT maintained adequate registration accuracy in breast tangential irradiation. These findings suggest its clinical applicability with appropriate acquisition angle settings.

PMID:
42839538
Bibliographic data and abstract were imported from PubMed on 07 Oct 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 16
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement