Clinical oral implants research. Oct 14, 2026. Epub Oct 14, 2026.
Abstract
To evaluate and compare the positional accuracy of complete arch intraoral digital implant scans obtained using a conventional scan body protocol and a customized metallic ring-based calibrated framework.
A completely edentulous arch restored with six implants was scanned using two complete arch intraoral digital impression protocols: a conventional scan body protocol (SB) and a metallic ring-based calibrated framework (CF). Twenty scans were obtained per group using the same intraoral scanner. A coordinate measuring machine (CMM) dataset served as the reference standard. All STL files were imported into dental CAD software to design a standardized virtual implant-supported bar, which was subsequently analyzed using reverse engineering software. Linear inter-implant distances and angular deviations were calculated. Trueness was defined as the absolute deviation from the CMM reference, and precision as the variability among repeated scans. Between-group comparisons were performed using independent t-tests with Welch correction and Holm adjustment (α = 0.05).
The mean linear deviation was 0.034 (0.004) mm for CF and 0.124 (0.019) mm for SB (p < 0.01). The mean angular deviation was 0.31 (0.05) degrees for CF and 0.55 (0.10) degrees for SB (p < 0.01).
The calibrated splinting protocol showed greater trueness and lower variability compared with the conventional scan body approach.
PMID:
42870725
Bibliographic data and abstract were imported from PubMed on 14 Oct 2026.
Read full publication at:
Please sign in
to see all details.
Advertisement
Stats
- Recommendations n/a n/a positive of 0 vote(s)
- Views 13
- Comments 0