Hiring in life sciences? Share your open positions with our professional community. Read more Close

Advertisement

Transient Ischemic Attack Due to Delamination of a Steerable Guiding Sheath: A Potential Complication During Atrial Fibrillation Ablation.

Created on 16 Oct 2026

Authors

Hafez Golzarian, Mohammad Khalil Kanjwal, Fayaz Hakim

Published in

The Journal of innovations in cardiac rhythm management. Volume 17. Issue 9. Pages 6854-6856. Epub Sep 15, 2026.

Abstract

We report a case of transient ischemic attack due to a thromboembolic event secondary to delamination of the steerable guiding sheath, CARTO VIZIGO™ (J&J MedTech, New Brunswick, NJ, USA), indicating a device deficiency in structural integrity. This case illustrates an unforeseen device-related complication, emphasizing the importance of heightened vigilance for this potential complication during catheter ablations incorporating this sheath.