A Novel Semi-Cannulated Screw with Cement Augmentation for Revision Lateral Mass Fixation: A Cyclic Loading Biomechanical Study.

Created on 16 Oct 2026

Authors

Canyang Huang, Zhiping Huang, Zhenqiang Wang, Xiuhua Wu, Qingan Zhu, Dongbin Qu

Published in

Acta orthopaedica Belgica. Volume 92. Issue 3. Oct 08, 2026. Epub Oct 08, 2026.

Abstract

Dongbin Qu, Division of Spine Surgery, Department of Orthopaedics, Nanfang Hospital, Southern Medical University, Guangzhou, Guangdong 510000, P. R. China. Email: [email protected] Canyang Huang, Division of Spine Surgery, Department of Orthopaedics, Quanzhou First Hospital Affiliated to Fujian Medical University, Quanzhou, Fujian 362000, P. R. China. Email: [email protected] Purpose: This study biomechanically compared a novel semi-cannulated lateral mass screw with cement augmentation against conventional solid screws for stabilizing loosened cervical lateral masses under cyclic loading.
Eight fresh human cervical specimens underwent μCT scanning for BMD assessment. Each specimen received both screw types: on the augmentation side, semi-cannulated screws were initially loosened via pull-out testing then reinserted with cement; on the (contralateral) primary side, conventional solid screws were placed Awithout augmentation. After cyclic loading, vertical displacement, toggling angle, and rotation center distance (ORC) at the screw-bone interface were measured, followed by final pull-out force tests.
The mean BMD of the lateral masses was 305.05 ± 48.33 mg/cm³. Following cyclic loading, the augmentation side exhibited a significantly higher pull-out force than its initial state (P < 0.01), though not statistically greater than the primary side. Under cyclic loading, the augmentation side demonstrated significantly superior stability: vertical displacement increased by only 6.25%, compared to a 70.69% increase on the primary side (P < 0.05). The toggling angle decreased by 2.36% on the augmentation side, versus a significant 79.92% increase on the primary side (P < 0.05). The ORC increased by 2.13% on the augmentation side, markedly lower than the 38.03% increase on the primary side.
The novel semi-cannulated lateral mass screw with cement augmentation demonstrated superior dynamic stability and significantly improved fixation strength after cyclic loading compared to conventional solid screws, offering a promising revision strategy for cervical spines with screw loosening or osteoporosis.

PMID:
42885449
Bibliographic data and abstract were imported from PubMed on 16 Oct 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 19
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement