Beyond Gamma Analysis: An Accessibility-oriented Framework for Comparative Evaluation of Patient-specific Radiotherapy Quality Assurance.

Created on 18 Oct 2026

Authors

Vikram, Raj Pal Singh, Manindra Bhushan

Published in

Journal of medical physics. Volume 51. Issue 3. Pages 637-645. Epub Sep 23, 2026.

Abstract

Patient-specific quality assurance (QA) in radiotherapy is traditionally performed using measurement-based systems that require dedicated hardware, machine time, and manpower. Although software-based QA approaches have emerged as efficient alternatives, their comparative accessibility and operational performance remain insufficiently investigated. This study evaluated conventional and software-based QA methodologies using integrated dosimetric, temporal, and economic metrics.
A total of 90 clinically approved treatment plans, including head-and-neck, pelvis, and lung volumetric modulated arc therapy/stereotactic body radiotherapy cases, were retrospectively analysed. All plans were generated in Eclipse TPS v15.6 and evaluated using ArcCHECK phantom-based QA, electronic portal imaging device (EPID)-based QA, and log file-based QA. Gamma analysis was performed using 3%/2 mm criteria with a 10% dose threshold. Workflow time and relative cost burden were recorded for each method. Three composite indices were defined: QA Cost Efficiency Index (QCEI), QA Efficiency Index (QEI), and QA Accessibility Index (QAI). Statistical comparisons were performed using paired analysis with Bonferroni correction.
All QA methods achieved clinically acceptable gamma passing rates. Log file-based QA demonstrated the highest median gamma passing rate; however, the absolute differences between methodologies were small, and all approaches exceeded the predefined clinical acceptance criterion (≥95%). ArcCHECK showed the greatest workflow time and cost burden, whereas log file-based QA demonstrated the highest QCEI, QEI, and QAI values. EPID-based QA showed intermediate performance. All pairwise differences in accessibility metrics were statistically significant (P < 0.05).
Software-based QA approaches significantly improved operational efficiency and accessibility while maintaining clinically acceptable dosimetric agreement. However, further prospective, multi-institutional validation of the proposed accessibility framework is warranted.

PMID:
42891454
Bibliographic data and abstract were imported from PubMed on 18 Oct 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 14
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement