Hiring in life sciences? Share your open positions with our professional community. Read more Close

Advertisement

New reconstruction algorithm for stack artifact reduction in breathing-adaptive four-dimensional computed tomography.

Created on 18 Oct 2026

Authors

Juliane Szkitsak, Maximilian Heider, Niklas Andre Lackner, Charlotte Frei, Allison Lamrani, Christopher Dürrbeck, Annika Oleszczuk, Mushawar Ahmad, Jannis Dickmann, Janis Langkrär, Oliver J Ott, Stefanie Corradini, Rainer Fietkau, Christoph Bert

Published in

Physics and imaging in radiation oncology. Volume 41. Pages 101102. Epub Oct 03, 2026.

Abstract

Even with new adaptive four-dimensional computed tomography (4DCT) techniques, artifacts can still occur with strong irregular breathing amplitudes. A new image reconstruction algorithm, currently available for cardiac CT, was tested on 4DCT data for its performance to reduce stack artifacts and thus improve anatomical correctness.
The 'ZeeFree' algorithm employs a deformable image registration during reconstruction to reduce remaining stack artifacts. Its performance was evaluated using data of 20 irregular breathing ('challenging group') and 10 randomly selected ('verification group') patients with tumors in the thorax. Using a 5-score-scale, image quality in terms of stack artifacts was rated by three physicians and three physicists. Statistical significance was assessed using the Wilcoxon signed-rank test for the mean ratings and the Chi-square test for anatomical correctness.
Expert scoring revealed that 88.3% of the amplitude-based ZeeFree reconstructions showed (almost) no artifacts (rated with ≥ 4) compared to 66.8% of the standard reconstruction. With phase-based binning, 70.0% of the ZeeFree and 23.3% of standard were rated with a score ≥ 4. Mean scores were 4.1 and 4.0 for amplitude- and phase-based binning with ZeeFree and 3.4 and 2.5 for standard reconstruction, respectively. Differences were significant (p < 0.001). 91.1% (ZeeFree) and 41.1% (standard) were rated as anatomically correct. Again, differences were significant (p < 0.001).
This study demonstrated a statistically significant reduction of stack artifacts, resulting in improved anatomical correctness when ZeeFree was applied to 4DCTs. The algorithm was not observed to introduce new artifacts on already artifact-free images.

PMID:
42892968
Bibliographic data and abstract were imported from PubMed on 18 Oct 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 5
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement