[Subject Thickness Measurement and Entrance Surface Dose Estimation Using an RGB-D Camera in Chest Radiography].

Created on 19 Oct 2026

Authors

Kenji Yasue, Atsushi Urikura, Koichi Hanada, Hiraku Fuse

Published in

Nihon Hoshasen Gijutsu Gakkai zasshi. Volume 82. Issue 11.

Abstract

This study aimed to quantify the effect of measurement duration on subject thickness accuracy using an RGB-D camera at a source-to-image distance (SID) of 200 cm for chest radiography and to evaluate the accuracy of entrance surface dose (ESD) estimation based on 1-s measurements.
A chest radiography phantom was evaluated at subject thicknesses of 20, 25, 30, and 35 cm. Subject thickness was measured five times at measurement durations of 1, 5, and 10 s and compared with caliper measurements. ESD estimated from the mean subject thickness obtained over 1 s was compared with the reference ESD calculated from the caliper-based thickness.
The maximum subject thickness error was 0.48 cm, the maximum relative error was +1.4%, and the maximum standard deviation was 0.12 cm. No significant differences among measurement durations were found at 20, 25, or 30 cm. At 35 cm, an overall difference was found, but no pairwise difference was significant after adjustment for multiple comparisons. The maximum relative error in ESD estimation was +0.6% at 35 cm.
Under static phantom conditions at an SID of 200 cm, subject thickness measurements obtained over 1 s and the resulting ESD estimates were close to the reference values. These findings support further evaluation of short-duration RGB-D measurements for chest radiography.

PMID:
42893864
Bibliographic data and abstract were imported from PubMed on 19 Oct 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 3
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement