Authors
Yue Yang, Matthieu Chabanas, Carrie Reale, Annie Benson, Jason Slagle, Matthew Weinger, Michael Topf, Jie Ying Wu
Published in
Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention. Volume 16893. Pages 23-31. Epub Sep 26, 2026.
Abstract
Positive margins are common in head and neck squamous cell carcinoma, yet intraoperative re-resection is often imprecise because margin locations are typically communicated verbally from pathology. We present an all-in-one augmented reality (AR) system that relocalizes positive margins from a resected specimen to the resection bed and visualizes them in situ using HoloLens 2 depth sensing and fully automated markerless surface registration. In a silicone phantom study with six medical trainees, markerless registration achieved target registration errors comparable to a marker-based baseline (median 1.8 mm vs. 1.7 mm; maximum < 4 mm). In a margin relocalization task, AR guidance reduced error from verbal guidance (median 14.2 mm) to a few millimeters (median 3.2 mm), with all AR localizations within 5mm error. These results support the feasibility of markerless AR margin guidance for more precise intraoperative re-resection.
PMID:
42894872
Bibliographic data and abstract were imported from PubMed on 19 Oct 2026.