Authors
Karl-Johan Fredén Jansson, Johannes Kocher, Martin L Johansson
Published in
Medical devices (Auckland, N.Z.). Volume 19. Pages 631032. Epub Oct 14, 2026.
Abstract
Piezoelectric transducers are gaining prominence in hearing implants due to their compatibility with magnetic resonance imaging, but commercial alternatives suffer from excessive power consumption. The objective of this study was to propose and evaluate a method of exclusively using passive electronic components in an implanted electronic circuit with low current consumption for driving of a piezo transducer-based hearing aid implant system.
The study sample was a hearing aid system of a prototype piezo transducer and a commercially available sound processor. Output force level and total harmonic distortion were measured by a skull simulator inside an anechoic test chamber at constant sound pressure levels of 70, 80 and 90 dB SPL. Current consumption measurements and a continuous sound exposure test determined the battery lifetime.
The current consumption was 5.6 mA (a battery lifetime of 7-9 days), the maximum total harmonic distortion was 2.15% at 70 dB SPL and maximum power output was 111.7 dB re. 1 µN at 90 dB SPL.
The findings of the passive design presented in this study indicate that piezo transducers can be driven by significantly lower current consumption than today's commercially available alternatives that are based on active electronics.
PMID:
42894782
Bibliographic data and abstract were imported from PubMed on 19 Oct 2026.