Authors
Min-Sheng Hung, Sheng-Ze Qiu, Yu-Cheng Wu
Published in
Bioanalysis. Pages 1-10. Oct 20, 2026. Epub Oct 20, 2026.
Abstract
Circulating tumor cells (CTCs), which detach from primary tumors and enter the peripheral blood, serve as biomarkers for monitoring a patient's status following cancer therapy.
In this study, we developed a microfluidic platform integrating dielectrophoresis and image recognition for the capture and detection of CTCs from blood sample. Experimental parameters included applied electric field strengths ranging from 12.5 to 32.5 × 104 V/m (at 1 MHz) and sample flow rates of 6.25-31.25 mm/min. Fluorescent antibodies were employed to label captured CTCs for real-time detection. Furthermore, a detection software utilizing OpenCV was developed to label CTCs through image processing. The identification performance of the self-developed software was evaluated and compared with machine learning algorithms.
Results demonstrated that the optimal CTC separation efficiency of 78 ± 2% was achieved at an electric field of 25 × 104 V/m and a flow rate of 18.75 mm/min. While the self-developed software achieved a lower recall than the YOLOv8 model (85 ± 4% vs. 92 ± 2%), its precision matched that of YOLOv8 (90 ± 3% vs. 89 ± 4%).
PMID:
42902582
Bibliographic data and abstract were imported from PubMed on 21 Oct 2026.
Read full publication at:
Please sign in
to see all details.
Advertisement
Stats
- Recommendations n/a n/a positive of 0 vote(s)
- Views 4
- Comments 0