Waived semen analyzers in clinical practice: analytical standards, regulatory categorization, and the limits of automated assessment.

Created on 27 Oct 2026

Authors

David H McCulloh, Jeff Boldt, Charles H Muller, Carol L Curchoe

Published in

Journal of assisted reproduction and genetics. Oct 26, 2026. Epub Oct 26, 2026.

Abstract

The award of Clinical Laboratory Improvement Amendment (CLIA)-waived status to automated semen analyzers, including the LensHooke (Bonraybio) and SQA-iOw (Medical Electronic Systems), has prompted substantive debate regarding analytic adequacy, regulatory classification, and appropriate clinical use.
To synthesize academic and industry perspectives on the analytical standards, regulatory categorization, and clinical deployment of waived automated semen analyzers, and to provide practical guidance on laboratory validation.
We reviewed peer-reviewed publications evaluating the SQA-V and SQA-iOw, the United States (US) Food and Drug Administration's (FDA's) 510(k) and CLIA waiver documentation, the US Centers for Medicaid and Medicare Services (CMS) CPT transmittals, the College of American Pathologists (CAP) Reproductive Lab Medicine Checklist requirements, and World Health Organization (WHO) Laboratory Manual guidance on computer-assisted semen analysis.
Waived semen analyzers offer genuine advantages in workflow efficiency and intra-instrument reproducibility and are appropriate for preliminary assessment of major male factor causes of infertility. Documented limitations in morphology accuracy near the 4% Tygerberg threshold, motility detection at low values, and handling of complex samples make them insufficient as stand-alone replacements for qualified manual analysis in high-stakes contexts including azoospermia determination and pharmacologic research as well as studies on infertility pathophysiology and diagnosis, and fertility prognosis. The scope of the CLIA waivers is settled; the applicable CPT billing codes remain unresolved and require professional society engagement with the American Medical Association's (AMA's) CPT Editorial Panel. Laboratory validation is a professional obligation regardless of CLIA complexity level and is also required internationally by ISO 15189.

PMID:
42932108
Bibliographic data and abstract were imported from PubMed on 27 Oct 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 6
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement