Authors
Federico Turreni, Pietro Francia, Maria Luisa Loricchio, Renzo Iulianella, Antonino Granatelli, Marco Testa
Published in
Pacing and clinical electrophysiology : PACE. Oct 27, 2026. Epub Oct 27, 2026.
Abstract
S-ICD efficacy depends on extracardiac shock vector geometry. We report late obesity-associated geometric deterioration after substantial weight gain. Rising shock impedance prompted reassessment at generator replacement, revealing pocket dehiscence, adipose interposition, and generator superficialization. Because fibrosis precluded parasternal lead mobilization by simple traction, sub-serratus generator repositioning improved the PRAETORIAN score and shock impedance. Despite a subsequent rise to 130 Ω, defibrillation succeeded with a 65 J shock. S-ICD geometry should be considered dynamic, particularly after major changes in body habitus.
PMID:
42944557
Bibliographic data and abstract were imported from PubMed on 27 Oct 2026.