Published in
Documenta ophthalmologica. Advances in ophthalmology. Oct 30, 2026. Epub Oct 30, 2026.
Abstract
To establish the baseline data of static and dynamic characteristics of the pupil light reflex in Indian eyes across a wide age range using Pupil N, a newly developed handheld, objective pupillometer and thereby compare the present outcomes with a comparable dataset in Indian eyes obtained using the NPi-200, the current standard of care in objective pupillometry.
Pupil N elicits the light reflex by delivering two 1-s-long visible light pulses of 125 lx intensity onto the eye with 4-s of interstimulus interval, all after initial 6-s of dark adaptation. Pupil diameters recorded using an infrared sensitive camera (30 fps) are processed using a commercial smartphone to derive six outcome variables: pre- and post-stimulated diameter, miosis (normalized to pre-stimulation values), constriction and dilation peak velocities and response latency. Baseline data were collected in this study in both eyes on 290 healthy participants (3-86 years of age) and the right eye data of adults were compared with those obtained using NPi-200 by Rattan et al. [37].
Baseline values calculated for each age decade showed younger eyes having larger pre-and post-stimulation diameters, faster constriction and dilation peak velocities, relative to older eyes (p < 0.001). The magnitude of miosis and response latency were similar across ages, except for larger miosis and shorter latencies in the first decade of life (p < 0.001). Data trends were similar to Rattan et al., but the absolute values differed across outcome variables (p < 0.001), except for pre-stimulation diameter (p = 0.06).
The baseline values of pupil light reflex reported here for Pupil N may be used as future reference standards for interpreting data of individuals with neurological dysfunction. The inter-device differences reported here call for caution while using these devices interchangeably for eliciting/interpreting the pupil light reflex in clinical and research environments.