Authors
Guyue Zhou, Wenchen Wang, Yiqing Yin, Haotian Miao, Zhan Liu, Shufeng Wei, Wenhui Yang, Yunxing Song, Feng Liu, Yaohui Wang
Published in
The Review of scientific instruments. Volume 97. Issue 11. Nov 01, 2026.
Abstract
High-quality magnetic resonance imaging (MRI) relies on accurate magnetic field characterization; however, commercial nuclear magnetic resonance (NMR) magnetic field probes are often limited in boundary-accessible field mapping for bore-constrained magnets. In this study, a miniature free-induction-decay-based NMR magnetic field probe was developed, integrating a solenoid microcoil, a compact tuning/matching network, and a liquid sample unit, with a maximum width of ∼8 mm. Compared with a commercial NMR reference probe, the developed microprobe showed close agreement in single-point validation and yielded a mean absolute deviation of 0.27 μT in axial comparison measurements (corresponding to 0.039 ppm at 7 T), confirming its field-measurement accuracy and spatial tracking capability. Temperature-sensitivity tests over a range of 15-35 °C showed that the repeated-measurement standard deviation remained within 0.019-0.046 ppm, indicating good measurement stability. In wall-adjacent measurements inside a 172 mm-bore magnet, the maximum accessible trajectory diameter increased from 152 to 164 mm, corresponding to a 7.9% improvement in an accessible mapping range. Spherical-surface mapping of a 50 mm diameter spherical volume further yields a mean absolute deviation of 0.143 μT and a Pearson correlation coefficient greater than 0.99. These results demonstrate that the proposed microprobe provides a low-cost, stable, high-precision, and broadly applicable solution for extending accessible magnetic field mapping in bore-constrained MRI magnets.
PMID:
42969823
Bibliographic data and abstract were imported from PubMed on 02 Nov 2026.
Read full publication at:
Please sign in
to see all details.
Advertisement
Stats
- Recommendations n/a n/a positive of 0 vote(s)
- Views 24
- Comments 0