[Finite element analysis of a novel anatomical locking plate in hyperextension varus tibial plateau fractures].

Created on 06 Nov 2026

Authors

Jiafan Zhang, Anhua Long, Dacheng Han, Zichao Jia, Yakui Zhang

Published in

Zhongguo gu shang = China journal of orthopaedics and traumatology. Volume 39. Issue 10. Pages 1037-43. Oct 25, 2026.

Abstract

To introduce a novel anatomic locking plate for managing (hyperextension varus tibial plateau fracture, HVTPF) and to analyze the biomechanical characteristics of the novel anatomical plate using finite element analysis (FEA).
A three-dimensional digital model of the full length of left tibia was obtained by CT scanning. A HVTPF model was then created. The acquired fracture model was assembled with the novel anatomic locking plate (NALP) and the conventional anatomic locking plate (CALP). The vertical displacement and stress of the anteromedial fragments were evaluated under axial loads of 500 N and 1000 N.
Under a 500 N load, the maximum displacement of the overall models in the NALP and CALP groups was 2.245 5 mm and 3.459 0 mm, respectively, while the maximum displacement of the anteromedial fracture fragment was 2.187 7 mm and 3.428 9 mm. The directional displacements of the anteromedial fracture fragment in the NALP group were -0.626 4 mm and 0.092 0 mm, compared to -1.000 4 mm and 0.149 7 mm in the CALP group. The maximum stresses of the overall models were 148.23 MPa and 355.44 MPa, respectively, and the maximum stresses of the anteromedial fracture fragment were 26.907 MPa and 84.706 MPa. Under a 1 000 N load, the maximum displacement of the overall models in the NALP and CALP groups was 4.483 3 mm and 6.856 6 mm, respectively, while the maximum displacement of the anteromedial fracture fragment was 4.374 3 mm and 6.796 6 mm. The directional displacements of the anteromedial fracture fragment in the NALP group were -1.247 2 mm and 0.189 8 mm, compared to -1.980 0 mm and 0.299 0 mm in the CALP group. The maximum stresses of the overall models were 293.46 MPa and 2 485 MPa, respectively, and the maximum stresses of the anteromedial fracture fragment were 50.586 MPa and 105.88 MPa.
Based on our FEA results, the NALP demonstrates superior stability and mechanical properties in managing HVTPF. There it could be a good option for treating HVTPF in further clinical applications.

PMID:
42988401
Bibliographic data and abstract were imported from PubMed on 06 Nov 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 10
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement