Abstract
We describe mechanical leaflet traversal as an alternative to electrosurgical traversal for BICORN during valve-in-valve transcatheter aortic valve replacement in a patient with unfavorable cardiac orientation and extreme bilateral coronary obstruction risk.
The ShortCut device created targeted basal splits in both coronary leaflets. The left coronary cusp split was crossed using a Pilot 100 guidewire (Abbott) and microcatheter with an EBU 3.5 guide catheter (Medtronic) and Agilis steerable introducer (Abbott) support. After balloon dilation, the left coronary cusp was translocated with an 18-mm True Dilatation Balloon (BD). A 20-mm SAPIEN 3 Ultra transcatheter heart valve (Edwards Lifesciences) was then advanced through the right coronary cusp split and deployed within the pre-existing 23-mm SAPIEN 3 valve (Edwards Lifesciences).
Mechanical traversal requires precise targeting. Leaflet translocation may cause embolization, severe aortic insufficiency, or hemodynamic compromise, requiring readiness for immediate valve deployment.
Mechanical traversal may facilitate BICORN when cardiac rotation, unfavorable catheter trajectory, neosinus malalignment, eccentric coronary takeoffs, or leaflet fibrocalcification make electrosurgical traversal challenging.