Abstract
While safe positive intrarenal pressure (IRP) thresholds during retrograde intrarenal surgery (RIRS) are well known, negative outlet wall pressure applied to Flexible and Navigable Suction Sheaths (FANS) remains uncharacterized.
To evaluate IRP dynamics under varying irrigation heights and negative suction pressures, providing foundational data to support evidence-based protocols in suction-assisted endourology.
A silicone kidney model was instrumented with an 11-13Fr/46 cm FANS and a 7.5Fr single-use flexible ureteroscope. Real-time IRP was measured via pressure wire during four sequential phases: irrigation (30 seconds), suction (30 seconds), simulated channel blockage (90 seconds), and unblocking (5 seconds). Four experimental groups evaluated combinations of different irrigation heights (40 and 100 cmH2O) and negative pressures (-100 and -375 mmHg). Each combination was replicated three times, yielding 12 independent experiments. The primary outcome was the rate of IRP change (mmHg/s) during unblocking, and the secondary outcome was the maximum IRP observed during blockage. Data were analyzed using one-way ANOVA with Tukey's post hoc test (p < 0.05).
During unblocking, -375 mmHg suction produced significantly faster IRP reduction than -100 mmHg across all subgroups (mean differences 7.6-10.45 mmHg/s, p < 0.05). During blockage, 100 cmH2O irrigation generated significantly higher peak pressures than 40 cmH2O (mean differences 7.97-18.37 mmHg, p < 0.05). The combination of high irrigation pressure and high negative suction resulted in the most extreme and rapid pressure fluctuations, with changes occurring within fractions of 10-milliseconds.
Negative outlet wall pressure significantly influences IRP dynamics during RIRS with FANS, producing rapid and extreme pressure changes. Pressure changes may occur on a timescale that could challenge manual recognition and corrective intervention. Standardized pressure monitoring and automated control systems might become a promising approach to ensuring patient safety during suction-assisted endourology.