Authors
Yuya Nakamura, Ayaka Oda, Haruka Miyazaki, Shuhei Arai, Taku Asano, Toshiro Shinke
Published in
JACC. Case reports. Pages 111205. Nov 12, 2026. Epub Nov 12, 2026.
Abstract
Dual-chamber leadless pacing depends on beat-to-beat implant-to-implant communication; heart failure-related geometric changes may expose limited communication reserve.
A 92-year-old woman with complete atrioventricular block developed heart failure 9 months after dual-chamber leadless pacemaker implantation, with reduced atrial-to-ventricular (A-to-V) throughput, atrioventricular dissociation, ventricular functional and diastolic mitral regurgitation, and reduced left ventricular ejection fraction. Decongestion was accompanied by shortening of the projected interdevice distance, improvement in A-to-V throughput and atrioventricular synchrony, and disappearance of diastolic mitral regurgitation. Heart failure recurred despite preserved throughput and continued medical therapy. Upgrade to a transvenous dual-chamber pacemaker with left bundle branch area pacing narrowed the QRS and improved left ventricular function and mitral regurgitation.
Pacing-induced cardiomyopathy appeared to be the dominant substrate, with dynamic A-to-V limitation as a reversible aggravating factor. Left bundle branch area pacing addressed both atrioventricular and ventricular dyssynchrony.
PMID:
43018525
Bibliographic data and abstract were imported from PubMed on 12 Nov 2026.