Authors
Alan-Shaun Wilkinson, Kate Walker, Laima Ozolina, Romana Bon, Andrew Wallace, Michael Charles Allwood
Published in
PloS one. Volume 21. Issue 11. Pages e0337729. Epub Nov 12, 2026.
Abstract
The 2016 Draft National Institute for Occupational Safety and Health (NIOSH) universal protocol "A Performance Test Protocol for Closed System Transfer Devices Used During Pharmacy Compounding and Administration of Hazardous Drugs." employing ATD-GC-MS with 2-Phenoxyethanol (2-POE, 2.5% w/v in water) as challenge agent (2016 Draft NIOSH universal protocol) was used to assess containment performance of two barrier and two air-cleaning CSTDs in a head-to-head test (n = 4) for both NIOSH tasks 1 (drug preparation) and 2 (drug administration). The study objective was to assess the containment performance of both physical barrier and air-cleaning CSTDs.
ChemoLock™ barrier (ICU Medical, USA), ChemoLock™ air-cleaning (ICU Medical, USA), PhaSeal™ barrier (Becton Dickinson, USA), and Tevadaptor™ air-cleaning (Simplivia, IL - now marketed as Chemfort™) CSTDs were evaluated using the 2016 Draft NIOSH universal test protocol for Task 1 (n = 4) and Task2 (n = 4). A 100 mL empty IV bag, accessed with a CH-10 ChemoClave™ bag spike was used with the IV administration set in Task 2 to avoid pressurisation of the occluded administration set during the simulated "IV push". Air samples were collected (GilAir pump) at 100mL/minute for 30 minutes on Tenax TA thermal desorption (TD) tubes (Markes, UK). 2-POE, a semi-volatile surrogate mimicking hazardous drugs was prepared at 2.5% w/v solution in MilliQ water. This was aliquoted (50mL) into drug vials and crimp sealed with a butyl rubber stopper (Adelphi, UK). 2-POE was analysed against a calibration curve (n = 8) in the range 0.6ng to 200ng using analytical methods published previously. Chromeleon v7.3 CDS (Thermo Scientific, UK) was used for analysis based on detection of SIM ions (M/Z: 77, 94, 138). Blank TD tubes (n = 85) were used to determine the experimental LOD and LLOQ. Negative control tests (surrogate of MilliQ water) were performed for each CSTD Task 1 (n = 1) and Task 2 (n = 1). An open "needle and syringe" positive control test was performed at 5 and 10 microlitre volumes in the NIOSH chamber to demonstrate system suitability.
Chamber cleaning and analysis performed pre-test, gave average blank values (n = 85) of 0.05 ± 0.01 ppbv (95% CI). Experimental LOD and LLOQ for the tests were 0.16 ± 0.01 ppbv and 0.41 ± 0.01 ppbv respectively. The positive control tests produced signals of 3.13 ppbv (n = 1) and 6.59 ppbv (n = 1) for 5 microlitre and 10 microlitre releases of 2-POE respectively and demonstrated system suitability to quantify ~ 650 nanolitre liquid release volume. (1) ChemoLock™ barrier, (2) ChemoLock™ air-cleaning, (3) BD PhaSeal™ barrier and (4) Tevadaptor™ air-cleaning CSTDs all generated
PMID:
43019316
Bibliographic data and abstract were imported from PubMed on 13 Nov 2026.