Empirical and semi-empirical methods to determine the kV exponent (n value) for diagnostic X-ray machines.

Created on 13 Nov 2026

Authors

Wellen Rukundo, Natharius Nimbashabira, Samuel Wattisa Maberi, Christine Edna Emong

Published in

Radiation protection dosimetry. Nov 13, 2026. Epub Nov 13, 2026.

Abstract

The air kerma, and beam quality measurements were performed on 43 diagnostic X-ray machines (25 fixed and 18 mobile) that met acceptance criteria for tube potential accuracy, half-value layer, and tube output. Measurements were conducted at tube potentials from 40 kV to 120 kV using a calibrated solid-state dosimetry system at a fixed focus-detector distance of 1 m, normalized output (mGy/mAs) and half-value layer were determined at each setting of the tube potential. Two methods (method 1 - empirical and method 2 - semi-empirical) were used to derive the exponent value. In method 1, a power-law fit of tube output versus tube potential yielded an average exponent of $n$ ≈ 2.334 ± 0.118 based on the group of X-ray machines with high-power frequency generators, consistent with values reported for diagnostic systems with standard filtration [1, 2]. In method 2, the $n$ was derived as a function of tube potential by incorporating the measured variation of half-value layer with tube potential, explicitly accounting for effective filtration. This approach showed that $n$ decreases from ~2.69 at 40 kV to 2.09 at 120 kV, demonstrating greater output sensitivity to tube potential changes at lower tube potentials. Therefore, the study introduces a semi-empirical method for estimating the exponent ($n$ value) that incorporates measured half-value layer variation to account for effective beam hardening, which is neglected in conventional empirical models. This provides a practical tool for characterizing the dependence of X-ray tube output on tube protocol optimization and equipment performance assessment, applicable for quality control in diagnostic X-ray machines.

PMID:
43023683
Bibliographic data and abstract were imported from PubMed on 13 Nov 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 7
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement