Development of RF Probes and Preamplifiers for MRI at Low Fields.

Created on 18 Nov 2026

Authors

Charlotte R Sappo, Jonathan B Martin, Millie A Barron, Selin Kandemir, Xinqiang Yan, John C Gore

Published in

NMR in biomedicine. Volume 39. Issue 12.

Abstract

Magnetic resonance imaging (MRI) systems operating at low magnetic fields ( B 0 ≲ 0 . 25  T) have experienced renewed interest due to the emergence of portable, point-of-care (POC) scanners, and new approaches to mitigating the intrinsically lower signal-to-noise ratios (SNR). The performance of radiofrequency (RF) components is a large factor in determining overall system sensitivity. In contrast to conventional MRI, at higher fields where SNR is typically sample-noise dominated, many low field (LF) systems operate in a coil-noise dominated regime, requiring careful optimization of RF coil design, conductor losses, and receiver electronics. This manuscript surveys RF coil technologies and associated hardware considerations for MRI systems operating below approximately 0.18T, within the range that we designate as low field in this manuscript (0-250 mT). We begin by revisiting early low-field imaging systems and the historical development of RF coil architectures for large-volume imaging. The fundamental noise mechanisms governing low-frequency MR detection are then examined, including Johnson noise in conductors, sample-induced losses, receiver noise figure, and electromagnetic interference (EMI). Design strategies for improving sensitivity are discussed, including conductor optimization (e.g., Litz wire), coil geometry selection such as solenoidal and saddle configurations, and methods for managing coil bandwidth and quality factor. Recent developments in transmit-receive switching, preamplifier design, EMI mitigation, and coil-preamplifier co-design are also reviewed. Finally, we highlight emerging directions for RF hardware development in low-field MRI, including RF coils for encoding, multi-element receive arrays for parallel imaging, and receiver architectures inspired by related modalities such as magnetic particle imaging and ultrasound. Continued advances in RF coil and receiver-chain design will be essential for improving sensitivity and enabling low-field MRI systems for clinical and research applications.

PMID:
43046515
Bibliographic data and abstract were imported from PubMed on 18 Nov 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 17
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement