Abstract
The objective of this study was to investigate a new vortex whistle (VW) designed to produce a whistle tone at higher peak flows and more closely replicate the flow profile observed in a typical spirometer-obtained forced vital capacity (FVC). Comparison to both a clinical standard spirometry system (Koko Sx1000, nSpire Health) and a previously reported benchmark VW design was conducted.
Measures of FVC and the forced expiratory volume in the first second (FEV1) were obtained for 53 young adults (n = 27 M, n = 26 F) between the ages of 18-30 years using the Koko Sx1000 spirometer, the previously reported benchmark VW (VWBench), and a newly designed high-flow VW (VWHF). For each subject, the largest FVC and largest FEV1 were obtained from the elicited trials for each device, with these values then used for the FEV1/FVC ratio. Linear regression analyses including the Sex variable were computed to assess the strength of the relationship and extent of residual standard errors between FVC and FEV1 measures obtained using different devices.
The VWHF was observed to correlate strongly with the Koko Sx1000 spirometer for FVC (r = 0.96, r² = 0.92) and showed substantial improvement vs. the VWBench. The VWHF was also observed to correlate significantly strongly with the standard for FEV1 (R = 0.84, R² = 0.70, including the Sex variable), whereas the VWBench did not correlate significantly with the standard for FEV1. While the VWHF produced a mean FEV1/FVC ratio that was not significantly different from that observed using the Koko Sx1000, the VWHF estimate of FEV1/FVC had a weak correlation with the Koko Sx1000 and relatively high predictive error.
Basic respiratory measures of FVC, FEV1, and the FEV1/FVC ratio can provide the voice clinician with objective measures regarding respiratory function underlying voice production that result in a more comprehensive voice evaluation and lead to referral for detailed pulmonary function workup. Though there are still limitations regarding improved accuracy of peak flow, FEV1, and FEV1/FVC measurements, the results of this study expand our understanding of the potential of a VW to provide low-cost measures that aid in the assessment of respiratory function in settings where traditional spirometry may be unavailable or unaffordable.