Hiring in life sciences? Share your open positions with our professional community. Read more Close

Advertisement

Reported adverse event with the TriClip device for tricuspid Transcatheter edge-to-edge repair: Analysis of the MAUDE database.

Created on 21 Nov 2026

Authors

Asa Phichaphop, Toby Rogers, Edward T Ha, Ori Waksman, Itsik Ben-Dor, Lowell F Satler, Ron Waksman

Published in

Cardiovascular revascularization medicine : including molecular interventions. Nov 10, 2026. Epub Nov 10, 2026.

Abstract

The TriClip system (Abbott Cardiovascular) received United States Food and Drug Administration (FDA) approval on April 2, 2024, becoming the first dedicated device for tricuspid transcatheter edge-to-edge repair (T-TEER). Pivotal trial data demonstrated clinically meaningful improvements in quality of life with a favorable safety profile. However, real-world adverse events data following broad clinical adoption remain limited.
We queried the FDA Manufacturer and User Facility Device Experience (MAUDE) database to characterize adverse events reported during the first year following FDA approval of the TriClip system (April 2, 2024-April 1, 2025).
A total of 723 adverse event reports from 634 patients were identified. Single leaflet device attachment (SLDA) was the most frequently reported adverse event (n = 313, 43.3%). SLDA was identified intra-procedurally in 136 cases (43.5%). The majority of SLDA events were managed with additional clip implantation (n = 195, 62.3%), while surgical management was uncommon (n = 7, 2.2%). Device malfunction was the second most frequent adverse event (n = 221, 30.6%), followed by leaflet injury (n = 96, 13.3%). Device embolization was infrequently reported (n = 10, 1.4%), with half successfully retrieved percutaneously. All-cause death was reported in four cases. The number of reported events increased from 326 in the first 6 months to 391 in the subsequent 6 months, consistent with growing clinical adoption, while SLDA remained the most common event across both intervals.
In the first year following FDA approval of TriClip, SLDA was the most common reported event and was predominantly managed with additional clip implantation, with surgical intervention rarely required. Because the MAUDE database lacks a procedural denominator, reports reflect the distribution of adverse event types rather than the true incidence rate.

PMID:
43064649
Bibliographic data and abstract were imported from PubMed on 21 Nov 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 14
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement