Optimisation of reconstruction parameters for lutetium-177 SPECT imaging: comparative performance of cadmium-zinc-telluride and NaI(Tl) detectors.

Created on 23 Nov 2026

Authors

Hongjin Zhou, Shen Wang, Wei Zheng, Ruyi Zhang, Qiang Jia

Published in

Nuclear medicine communications. Nov 23, 2026. Epub Nov 23, 2026.

Abstract

Lutetium-177 (177Lu) is a promising theranostic radionuclide. Accurate quantitative SPECT imaging is essential for 177Lu patient-specific dosimetry. Although the 113 keV photopeak has a higher scatter fraction, it matches the high energy resolution of cadmium-zinc-telluride (CZT) detectors and avoids 208 keV limitations. However, whether 113 keV-only imaging provides sufficient lesion detectability remains unclear. Focusing on the 113 keV photopeak, this study compared two dual-head SPECT/computed tomography systems: conventional NaI(Tl) scintillators with low-energy high-resolution collimators, and CZT semiconductor detectors with wide-energy high-resolution collimators. This study compares these technologies and establishes optimized reconstruction protocols.
National Electrical Manufacturers Association/International Electrotechnical Commission phantoms were scanned on both systems using the 113 keV photopeak. Ordered subset expectation maximization reconstructions were performed (2-12 iterations) with Gaussian [full-width-at-half-maximum (FWHM): 1.5-4.0 mm) or Butterworth filters (cut-off frequency: 0.25-0.50 cycle/pixel). Attenuation, scatter, and resolution corrections were applied. Image quality was quantified using background coefficient of variation (COV_B), contrast recovery coefficient (CRC), contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR).
CZT delivered superior image quality, achieving higher CRC, CNR, and SNR with lower COV_B than NaI(Tl). Optimal parameters were 2 iterations, 10 subsets, Gaussian FWHM 3.0 mm, and Butterworth cut-off 0.35 cycles/pixel (CZT) and 0.30 cycles/pixel [NaI(Tl)]. These settings yielded robust performance across target-to-background ratios of 8 : 1, 16 : 1, and 32 : 1.
For the 177Lu 113 keV peak, CZT detectors markedly enhance SPECT imaging, improving lesion detectability and quantification. Tailored reconstruction optimization further strengthens image quality and reliability, supporting precision dosimetry and CZT adoption in radionuclide therapy.

PMID:
43070011
Bibliographic data and abstract were imported from PubMed on 23 Nov 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 41
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement