Validity of a Consumer-Grade Bioelectrical Impedance Device for Body Composition Assessment Compared with a Three-Compartment Model in Recreationally Active Adults.

Created on 25 Nov 2026

Authors

Patrick G Saracino, Hannah K Eberhardt, Cecilia D Salas, Bernard Omolo

Published in

Applied physiology, nutrition, and metabolism = Physiologie appliquee, nutrition et metabolisme. Nov 24, 2026. Epub Nov 24, 2026.

Abstract

This study aimed to validate a consumer multi-frequency bioelectrical impedance device against a 3-compartment (3C) model in recreationally active adults. Following an overnight fast and abstention from caffeine (12 h), alcohol (24 h), and strenuous exercise (24 h), participants underwent body composition assessment using a consumer multi-frequency bioelectrical impedance device (BIAH30; InBody Dial H30). A criterion 3C model was derived from body mass, body volume via air displacement plethysmography, and total body water via a validated multi-frequency bioelectrical impedance device. Fifty-five recreationally active participants (29M/26F; 22±4 years, 170.5±8.2 cm; 70.6±10.5 kg; 24.2±2.6 kg·m⁻²) completed the study. BF% (21.7±7.8% vs 19.9±7.8%) and FM (15.3±5.8 vs 14.0±5.8 kg) were higher with BIAH30 compared to 3C (p < 0.001), whereas FFM (55.5±10.6 vs 56.7±10.4 kg) was lower (p < 0.001). Bland-Altman analysis indicated mean biases of 1.9% (95% LOA: -4.0 to 7.7%), 1.3 kg (95% LOA: -2.7 to 5.3 kg), and -1.2 kg (95% LOA: -5.1 to 2.8 kg) for BF%, FM, and FFM, respectively, with no proportional bias. Lin's concordance correlation coefficient indicated moderate to strong agreement (CCC: 0.90-0.98). In conclusion, the InBody H30 demonstrated significant differences in BF%, FM, and FFM relative to 3C. Mean bias was small, no proportional bias was observed, and agreement was moderate to strong. However, 95% limits of agreement were wide, indicating notable individual variability. While the H30 may serve as a practical, low-cost option for group-level estimates or longitudinal tracking, it should not be considered interchangeable with multi-compartment models for precise individual-level measurement.

PMID:
43077687
Bibliographic data and abstract were imported from PubMed on 25 Nov 2026.

Read full publication at:
Please sign in to see all details.

Advertisement

Stats

  • Community rating n/a 0 votes
  • Reviewers' rating n/a 0 votes
  • Your rating

1-terrible, 9-excellent. How would you rate this publication? Sign in in to submit your rating.

  • Recommendations n/a n/a positive of 0 vote(s)
  • Views 15
  • Comments 0

Recommended by

  • No recommendations yet.

Post a comment

You need to be signed in to post comments. You can sign in here.

Comments

There are no comments yet.

Advertisement