Authors
Jana Dornakova, Thomas Bachman, Radim Brabec, Ekaterina Orlova, Jan Janota, Jakub Rafl
Published in
Frontiers in pediatrics. Volume 14. Pages 1971581. Epub Nov 10, 2026.
Abstract
The study aimed to determine what effect small changes in the set target range of an automated FiO2 system would have on SpO2 exposure.
Single center randomized crossover design, with auto SpO2 targets switched between 3 settings (standard, narrow and higher) every 12 h. VLBW infants needing extended respiratory support were studied. Primary outcomes were time in the intended range (88%-93% < 29 weeks PMA, otherwise 90%-95%) and time at SpO2 extremes (≤80%, >98% with supplemental oxygen). Additional levels of SpO2 above and below the target range were also evaluated.
Ten infants, birth weight 820 grams (IQR 673-943), were evaluated. At entry they were 14 days old (IQR 13-18) and participated for 9-30 days. Four hundred and thirty-one 12-hour epochs were evaluated. The oxygen demand was relatively low [FiO2 0.24 (IQR 0.22-0.30)] and oxygen saturation well controlled. They spent a median of 80% (IQR 68-87) of the time in the intended target range, with differences among the three settings not statistically significant. In contrast the % time with SpO2 ≤ 80% during the higher setting was reduced and the % time with SpO2 > 98% with the higher setting was increased. Narrow compared to the standard SpO2 exposures were not different. The descriptive levels of hyperoxemia and hypoxemia were consistent with the two primary SpO2 extremes.
Narrowing the target range had little impact, however shifting the target range up just one point had a marked impact of decreasing hypoxemia and increasing hyperoxemia.
ClinicalTrials.gov, identifier NCT06207994.
PMID:
43080582
Bibliographic data and abstract were imported from PubMed on 25 Nov 2026.