Authors
Mark L Munoz, Rittik Chaudhuri, Matthew J Eckert, Erik D Mueller
Published in
Military medicine. Nov 26, 2026. Epub Nov 26, 2026.
Abstract
Reliable oxygen supply is a major challenge during prolonged casualty care and long-range aeromedical transport. Cylinders provide high-concentration oxygen but impose finite endurance, weight, high-pressure handling, and resupply requirements; concentrators can generate oxygen while power is available. To our knowledge, no prior peer-reviewed study has directly compared concentrator and cylinder oxygen during prolonged mechanical ventilation in actual fixed-wing flight.
Four flight exposures on a Dassault Falcon 50 and Bombardier Challenger 650 tested 4 complete concentrator-ventilator configurations. Oxygen flow 1-5 L/min was crossed with minute ventilation 5.0-15.0 L/min, yielding 100 matched active cells. Falcon source grids were paired across flights using ventilator-outlet readings; Challenger grids were paired within its 2-flight campaign using test-lung-entrance readings. Because the complete prespecified grid was observed, primary analyses were descriptive. We calculated signed and absolute source-difference summaries and displayed the complete grid. Exploratory factorial ordinary least squares models used categorical configuration, flow, minute ventilation, and flow-by-minute-ventilation terms with HC3 standard errors. The study evaluated equipment only and involved no humans or animals; institutional review board and animal care review were not applicable.
The mean cylinder-minus-concentrator difference was 4.72 percentage points (pp; SD, 4.60); the concentrator was lower in 97 of 100 cells. The median absolute difference was 3.15 pp, and the 90th- and 95th-percentile absolute differences were 10.33 and 12.19 pp. Configuration means ranged from 1.39 to 7.96 pp. Sensitivity-analysis means ranged from 4.72 to 4.85 pp. Both aircraft powered the concentrator at 5 L/min directly from cabin outlets without modification or observed circuit interruption. Exploratory modeling identified configuration-associated differences. Tests of flow at minute ventilation 10 L/min, minute ventilation at flow 1 L/min, and the flow-by-minute-ventilation interaction were not significant.
Concentrator readings were generally lower; the median absolute difference was 3.15 pp, while upper-tail differences and configuration dependence preclude blanket interchangeability. The findings support configuration-specific evaluation of concentrators as cylinder extenders or candidate substitutes and evaluation of hybrid strategies in which compressed oxygen remains available for higher-demand conditions or contingencies. The study does not establish patient-level equivalence, isolated aircraft or ventilator effects, or universal replacement.
PMID:
43097693
Bibliographic data and abstract were imported from PubMed on 27 Nov 2026.